Laboratory & Scientific Equipments » STABILITY TEST CHAMBER Back

STABILITY TEST CHAMBER

    Name of Equipment

LABTOP® Micro controller (PID) based Stability Test Chamber

    Air Circulation

Motor and blower arrangement to have uniformity of Temperature under loaded condition. ( 2 blowers for 400 Liters and above)

    Alarm

Audio Visual Alarm if the temperature deviates from the preset temperature

    Trays

Stainless Steel trays.

    Lighting

Interior illumination for working area.

    Insulation

3” Thick CFC free PUF Insulation.

    Heating System

U shaped S.S. Nichrome wire air heaters.

    Observation Door

Inside see through unbreakable acrylic door

    Refrigeration System

Hermetically Sealed CFC Free Emerson Compressor with R 134a refrigerant having state of art by pass system for continues run of compressor.

    Humidity System

Non condensing type steam injection system with water level arrangement with insulation to save heat energy.

    Control System

Precise control of temperature and humidity by using CE approved micro controller based PID controller (Auto tuning type)

    Validation Port

50 mm validation Port with silicon rubber seal to insert sensor for validation purpose.

    Power

Works on 230 V AC single phase 50Hz

    Other

Castor wheels, MCB, Adjustable Tray Height arrangements. Heavy duty latch with lock & key

    Safety Features

1. Float switch to cutoff the mains supply in case of low water level in the boiler tank.2. Dedicated safety controller with separate sensor to cut off the supply in case of overshoot and undershoot of Temperature giving audio visual alarm.3. Additional safety thermostat to cut off the boiler system in case of overshoot of temperature.4. 180 Sec. compressor ‘ON’ delay timer to safe guard the compressor.

    Certifications

ISO 9001: 2008 , ISO 13485 : 2012 and CE

    Documentations

LABTOP will provide IQ, OQ, PQ documents with material test certificates and calibration reports

SPECIFICATION

   Temperature Range

20 to 60° C

   Temperature Accuracy

± 0.2° C

   Temperature Uniformity

2° C

    Humidity Range

40 to 98% RH.

   Humidity Accuracy

± 2% RH.

   Humidity Uniformity

± 3% RH.

   Alarm

Audio-Visual Alarm.