Laboratory & Scientific Equipments » Walk In Plant Growth Chamber Back
Name of Equipment |
LABTOP PLC controlled (PID) based Plant Growth Chamber |
|
Control System |
PLC with PID Controller (Auto tune), Safety controller, auto changeover of Standby systems & other event Managements with 3.8” color HMI with touch screen, USB printing port and Ethernet connectivity |
|
Air Circulation |
Two chamber blowers to have uniformity of Temperature under loaded condition. (Above 400Liters) |
|
Alarm |
Audio Visual Alarm if the temperature deviates from the preset temperature |
|
Trays |
Stainless Steel perforated trays. |
|
Insulation |
3” Thick CFC free PUF Insulation. |
|
Heating System |
U shaped S.S. Nichrome wire air heaters. |
|
Observation window |
Inside see through unbreakable acrylic door |
|
Refrigeration System |
Hermetically Sealed CFC Free Danfoss/Emerson Compressor with R 134a refrigerant having state of art By pass system for continues run of compressor. |
|
Humidity System |
Non condensing type steam injection system with water level arrangement with insulation to save heat energy. |
|
Validation Port |
50 mm validation Port Hole with silicon rubber seal to insert sensors for validation purpose. |
|
Power |
Works on 230 V AC single phase 50Hz |
|
Other |
Castor wheels, MCB, Adjustable Tray Height arrangement. Heavy duty latch with lock & key. |
|
Safety Features |
1. Float switch to cutoff the mains supply in case of low water level in the boiler tank. |
SPECIFICATION
Temperature Range |
10° C to 60° C |
Temperature Accuracy |
± 0.2° C |
Temperature Uniformity |
± 2° C |
Humidity Range |
40% RH to 95% RH. |
Humidity Accuracy |
± 2% RH. |
Humidity Uniformity |
± 3% RH. |